This HCI4 laboratory testing checklist guides technicians through commissioning and validating a lab scanning system. It covers mechanical checks (bolts, rams in sync), safety devices (E-stops, reset buttons), electrical verification (power supplies, limits, PLC inputs), and servo drive programming. Set and verify X, Y, and Z home and limit positions, check air pressure integrity, and run PuTTY commands for homing and light sensor tests. Record serial numbers for the laser profiler, linescan camera, spectrometer, and controllers. Confirm spectrometer calibration and illuminator alignment, deploy the correct Arduino program (250 um or 500 um), and log performance, hours, trays scanned, and photos.
With SafetyCulture you can
With SafetyCulture you can
About author
This community page makes available free workplace checklists and templates created by other users within the SafetyCulture community. SafetyCulture has re-published this content and where possible, has credited the original author. SafetyCulture has not verified the accuracy, reliability or suitability of any community content. You agree that your use of any of this content is in accordance with SafetyCulture’s Terms and Conditions.